Commands Listed in Alphabetical Order
MEASUrement:
MEAS<x>? (Query Only)
Returns all measurement parameters for the specied active measurement <x>.
Group
Measurement
Syntax
MEASUrement:MEAS<x>?
MEASUrement:MEAS<x>:COUNt? (Query O nly)
Returns the number of values accumulated for this measurement since the last
statistical reset. Values may be ignored if they generated an error. Measurements
are specied by x, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:COUNt?
Examples
MEASUREMENT:MEAS3:COUNT? might return :MEASUREMENT:MEAS3:C OUNT
3247.
MEASUrement:MEAS<x>:DELay? (Query Only)
Returns the delay measurement parameters for the measurement specied by
<x>, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:DELay?
Examples
MEASUREMENT:MEAS1? might return
:MEASUREMENT:MEAS1:DELAY:D IRECTION FORWARDS;EDGE1 RISE;EDGE2
RISE
.
MSO4000 and DPO4000 Series Programmer Manual 2-217