Commands Listed in Alphabetical Order
Syntax
CALibrate:RES
ults:FACtory?
CALibrate:RESults:SPC? (Query Only)
Returns the status of the SPC operation. This query does not initiate a SPC.
Group
Calibration and Diagnostic
Syntax
CALibrate
:RESults:SPC?
Related Commands
*CAL?
Returns
INIT indicates that SPC has never successfully completed.
PASS indicates that the last SPC operation passed.
FAIL indicates that the last SPC operation failed.
RUNNING indicates that the SPC operation is running.
Examples
CALIBRATE:RESULTS:SPC? returns the results of the last SPC operation: either
PASS or FAIL.
CH<x>? (Query Only)
Returns the vertical parameters for channel <x>, where x is the channel number.
Group
Vertical
Syntax
CH<x>?