Tektronix 784A User Manual Page 151

  • Download
  • Add to my manuals
  • Print
  • Page
    / 294
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 150
Performance Tests
TDS 684A, TDS 744A, & TDS 784A Service Manual
4–75
PAL Signal
Source
Digitizing Oscilloscope
75 Cable
75
Terminator
Output
Figure 4–34: Line Count Accuracy Test Hookup
H Press the main-menu Trigger Position.
H Press the side-menu to Set to 50%.
H Press the main-menu to Horiz Pos.
H Press the side-menu to Set to 50%.
H Use the HORIZONTAL POSITION knob to move the falling edge
of the sync pulse to two divisions to the left of center screen. See
Figure 4–35.
Page view 150
1 2 ... 146 147 148 149 150 151 152 153 154 155 156 ... 293 294

Comments to this Manuals

No comments