Tektronix 784A User Manual Page 102

  • Download
  • Add to my manuals
  • Print
  • Page
    / 294
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 101
Performance Tests
4–26
TDS 684A, TDS 744A, & TDS 784A Service Manual
TDS 7XXA Test Record (Cont.)
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
TDS 7XXA Performance Test MaximumOutgoingIncomingMinimum
CH 3 SIGNAL OUTPUT, 1 M Pk-Pk 88 mV __________ __________ Pk-Pk 132 mV
CH 3 SIGNAL OUTPUT, 50 Pk-Pk 44 mV __________ __________ Pk-Pk 66 mV
Probe Compensator Output Signal
Frequency (CH1 Freq.) 950 Hz __________ __________ 1.050 kHz
Voltage (difference) 495 mV __________ __________ 505 mV
Page view 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 293 294

Comments to this Manuals