Tektronix CPS250 User Manual Page 31

  • Download
  • Add to my manuals
  • Print
  • Page
    / 34
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 30
Appendix B: Maintenance
CPS250 User Manual
21
POWER Light ON, but No Output from A or B Supplies with Power
Supply connected to Circuit
1. Disconnect the CPS250 outputs from the circuit being tested.
Check A and B VOLTAGE and CURRENT controls.
2. If the outputs are good, check the circuit under test for a short or
low resistance.
POWER Light ON, but No Output from Any Supplies with Power Supply
Not Connected to Circuit
Contact your nearest Tektronix service center for servicing.
Page view 30
1 2 ... 26 27 28 29 30 31 32 33 34

Comments to this Manuals

No comments