Page view 33
Performance Tests
Performance Verification Procedures
TDS 500A Test Record (Cont.)
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
Performance Test MaximumOutgoingIncomingMinimum
Probe Compensator Output Signal
Frequency (CH1 Freq.) 950 Hz __________ __________ 1050 Hz
Voltage (difference) 495 mV __________ __________ 505 mV
Page view 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 113 114