Tektronix TDS3000 User Manual Page 60

  • Download
  • Add to my manuals
  • Print
  • Page
    / 140
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 59
Performance Verification
4–4
TDS3000 Series Service Manual
Performance Verification Procedures
Before beginning these procedures, two conditions must first be met:
The oscilloscope must have been operating continuously for ten minutes in
an environment the meets the operating range specifications for temperature
and humidity.
You must perform the Compensate Signal Path operation described on page
2–21. If the operating temperature changes by more than 10° C, you must
perform the Compensate Signal Path operation again.
The time required to complete the entire procedure is approximately one hour.
WARNING. Some procedures use hazardous voltages. To prevent electrical shock,
always set voltage source outputs to 0 V before making or changing any
interconnections.
This procedure uses internal routines to verify that the oscilloscope functions and
passes its internal self tests. No test equipment or hookups are required. Start the
self test with these steps:
1. Disconnect all probes and cables from the oscilloscope inputs.
2. Push the UTILITY menu button.
3. Push the System screen button to select Diags.
4. Push the Loop screen button and choose Once.
5. Push the Execute screen button.
6. Push the OK Confirm Run Test screen button.
A dialog box displays the result when the self test completes. Push the MENU
OFF screen button to continue operation.
Self Test
Page view 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 139 140

Comments to this Manuals

No comments